OperationDefinition for appending retina AI results to an existing DiagnosticReport.
Append results from AI analysis of retina images to a existing DiagnosticReport
The $append-retina-ai-result operation is used to add AI analysis results to an existing DiagnosticReport. This operation is invoked bye the regional integration platform after the AI system has analyzed retinal images.
See RetinaAppendAIResultOperation-Example for a an example of the parameters.
Add AI result to an examination using the following operation:
[baseUrl]/DiagnosticReport/[id]/$append-retina-ai-result
[id] is the identifier UUID (Universally Unique Identifier), sometimes referred to as a GUID, of the DiagnosticReport, a.k.a examination, to update.
URL: [base]/DiagnosticReport/[id]/$append-retina-ai-result
| Use | Name | Scope | Cardinality | Type | Binding | Documentation |
| IN | sectraStudyId | 1..1 | string | Sectra study ID for the retinal images that were analyzed. | ||
| IN | conclusion | 1..1 | CodeableConcept | Conclusion indicating the next step. MUST be a single code from the 1000 series. If the validation parameter is set to true, the next step will always be manual grading. | ||
| IN | daysUntilNextExamination | 0..1 | integer | Number of days until next examination. Only included if the conclusion code indicates that the next step is a new examination. | ||
| IN | leftDiabeticRetinopathy | 0..1 | decimal | Left eye diabetic retinopathy evaluation, a decimal between 0.0 and 5.0 | ||
| IN | leftDiabeticMacularEdema | 0..1 | boolean | Left eye diabetic macular edema evaluation. True if diabetic macular edema is present. | ||
| IN | rightDiabeticRetinopathy | 0..1 | decimal | Right eye diabetic retinopathy evaluation, a decimal between 0.0 and 5.0 | ||
| IN | rightDiabeticMacularEdema | 0..1 | boolean | Right eye diabetic macular edema evaluation. True if diabetic macular edema is present. | ||
| IN | rightEyeImage | 0..* | Observation (Retina Image Parameter) | Right eye images used in the AI analysis. MUST conform to RetinaImageParameter profile. | ||
| IN | leftEyeImage | 0..* | Observation (Retina Image Parameter) | Left eye images used in the AI analysis. MUST conform to RetinaImageParameter profile. | ||
| IN | aiDevice | 1..1 | Device (Retina AI Device) | AI Device that performed the analysis. Includes AI product name, algorithm version and protocol version. | ||
| IN | fullReport | 1..1 | Attachment | Technical details for further reference about the AI grading. | ||
| IN | validation | 1..1 | boolean | Set true if the AI result is for validation purposes only and will not be used in clinical decision making. |